The MegaEISâ„¢ ME-2SP200 is a unique high-power hybrid workstation that integrates a precision Potentiostat Core (PTC) module into the robust MegaEIS architecture.
Engineered for advanced single-cell and short-stack research, it delivers up to 200 A of stable current. Crucially, its dual-mode operation provides a true bipolar ±5 V range in PTC mode alongside standard 0...5 V source capabilities, enabling highly sensitive zero-crossing electrochemical techniques traditionally limited to low-power laboratory instruments. This allows researchers to execute precision cyclic voltammetry, chronoamperometry, and true parallel EIS on large-format cells under heavy operational loads without compromising signal integrity.
See Unboxing, introduction to features, and practical live measurement demonstration of MegaEIS System.
High-Current Cyclic Voltammetry: Execute advanced electrochemical characterization on large-format industrial cells. The integrated PTC module enables precise cyclic voltammetry and linear sweeps at up to 200 A, allowing researchers to analyze catalyst activity and redox reactions on massive active areas that standard lab potentiostats cannot drive.
Large-Format Cell Validation: Bridge the gap between fundamental material science and industrial application. Drive single large-format cells or small short stacks natively with up to 200 A while capturing artifact-free, high-resolution EIS data to pinpoint membrane degradation or mass transport limitations under true operational stress.
Complex Electrochemical Sequencing: Automate demanding test protocols securely. Seamlessly link standard high-power load profiling with advanced potentiostatic methods—such as double-step chronoamperometry or custom polarization sequences—capturing transient responses with the strict precision required for advanced R&D.
| Power supply | 110 … 240 V AC / 50 … 60 Hz |
| Dimensions | Modular deign for 19“ rack cabinet |
| Protection rating | IP20 |
| Input voltage for PTC mode | ± 5 V |
| Input voltage for Source mode | 0 … 5 V |
| Total input current Total input current (internal + external load) |
Up to 200 A |
| Internal load current | 0 … 200 A |
| Max. internal power dissipation | 1 kW |
| Cooling | Water |
| Sampling | 24-bit ADCs, low-noise 50/60 Hz filtered sampling
Up to 1.25 Msps for EIS measurements |
| Measurement resolution | 0.001% of selected range |
| Accuracy | Voltage: ≤ 0.1% of range + 0.1% of reading
Current: ≤ 0.1% of range + 0.5% of reading |
| Acquitions methods | Constant V, I, Open circuit, Manual control, Chronoamperometry, Chronopotentiometry, Linear sweeps, Polarization curves, Current interruption, Load cycling/profiling, EIS – electrochemical impedance spectroscopy, Programmable sequences of all available methods |
| EIS frequency | 1 mHz … 100 kHz |
| EIS amplitude in PTC mode | 1 … 1000 mV, up to 50 A for < 1 kHz |
| EIS amplitude in Source mode | Up to 20% of maximum internal current for < 1 kHz |
| EIS channels | 2 |
| EIS Probes | ± 5 V |
| Connection | USB 2.0 |
| Software | Control software for MS Windows
Features: measurement setup and control, data acquisition, processing and visualization, pascal or python scripting, remote control by TCP/IP server for integration with top-level control system, examples for remote control using Python or LabVIEW. |
If these variants do not suit you, please contact us, we offer customized solutions!
To run your experiments, you’ll use the application CVM Manager – our measurement and control app that manages everything from workstation setup to real-time data acquisition and visualization. Just connect the MegaEIS System, choose your method, and let the science begin.
High-Power Fuel Cell & Electrolyzer Testing
Want to learn more about the solution, use cases, and references related to fuel-cell or electrolyzer testing? In our portfolio we have a solution for all scenarios including SOFC, etc.
High-Power Battery & Packs Testing
Want to learn more about the solution, use cases, and references related to battery-oriented testing? In our portfolio we have a solution for any type of batteries including redox-flow, etc.